Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Beta 230: (1 year) CAL VER /5500 DATE: 02-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 41 NUMBER OF LINES: 305 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to AC power and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 30min - REQUIRED 1.004 DISP Temperature: 25 +/-10 degC 1.004 DISP Humidity: 55 +/-10% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {VOLTAGE} 1.007 DISP Connect the 5500A to the UUT as follows: 1.007 DISP [32] 5500A NORMAL HI to UUT + V INPUT 1.007 DISP [32] 5500A NORMAL LO to UUT - V INPUT 1.008 DISP Press button to select input function. 1.008 DISP Press button to select Voltage. 1.009 5500 0.1 0.090000V 0.01% 1e-6U 2W 2.001 5500 1 0.90000V 0.01% 1e-5U 2W 3.001 5500 10 9.0000V 0.01% 1e-4U 2W 4.001 HEAD {mV} 4.002 DISP Press button to select input function. 4.002 DISP Press button to select millivolts. 4.003 5500 100 90.000mV 0.01% 0.001U 2W 5.001 5500 1000 900.00mV 0.01% 0.01U 2W 6.001 5500 10000 9000.0mV 0.01% 0.1U 2W 7.001 HEAD {CURRENT} 7.002 DISP Connect the 5500A to the UUT as follows: 7.002 DISP [32] 5500A AUX HI to UUT + mA INPUT 7.002 DISP [32] 5500A AUX LO to UUT - mA INPUT 7.003 DISP Press button to select input function. 7.003 DISP Press button to select current in mA. 7.004 5500 20 20.000mA 0.015% 0.001U 2W 8.001 5500 50 50.00mA 0.015% 0.01U 2W 9.001 HEAD {RESISTANCE} 9.002 DISP Connect the 5500A to the UUT as follows: 9.002 DISP [32] 5500A AUX HI to UUT 4W HI INPUT 9.002 DISP [32] 5500A AUX LO to UUT 4W LO INPUT 9.002 DISP [32] 5500A NORMAL HI to UUT 2W HI INPUT 9.002 DISP [32] 5500A NORMAL LO to UUT 2W LO INPUT 9.003 DISP Press button to select input function. 9.003 DISP Press button to select Ohms function. 9.003 DISP Select 4W termination operation and press . 9.004 5500 900 400.00Z 0.025/ 0.01U 4W 10.001 5500 900 900.00Z 0.05/ 0.01U 4W 11.001 HEAD {100ohm RTD(385)} 11.002 HEAD 100ohm RTD(385) - {INPUT(read)} 11.003 DISP Press to select an input function. 11.003 DISP Press to select the RTD function. 11.003 DISP Select the 385 type, and press . 11.003 DISP Select degrees Celsius, and press . 11.003 DISP Select 4W termination type, and press . 11.004 5500 800.00degC 0.26U R1 4W 12.001 HEAD {TEMPERATURE (TC)} 12.002 DISP This procedure will test one of the following 12.002 DISP thermocouple types: J,K,T,E,R,S,B, or N. # C,D, and G type TC not verified 5500A Output N/A). 12.003 MEM2 Which thermocouple type do you wish to test (JKTERSBN)? 12.004 JMPL TYPE_J ZCMPI(MEM2, "J") 12.005 JMPL TYPE_K ZCMPI(MEM2, "K") 12.006 JMPL TYPE_T ZCMPI(MEM2, "T") 12.007 JMPL TYPE_E ZCMPI(MEM2, "E") 12.008 JMPL TYPE_E ZCMPI(MEM2, "R") 12.009 JMPL TYPE_E ZCMPI(MEM2, "S") 12.010 JMPL TYPE_E ZCMPI(MEM2, "B") 12.011 JMPL TYPE_E ZCMPI(MEM2, "N") 12.012 DISP Only the following thermocouples are supported 12.012 DISP by this calibrator: J,K,T,E,R,S,B,N. 12.013 JMP 12.003 12.014 LABEL TYPE_T 12.015 HEAD {T-type THERMOCOUPLE} 12.016 DISP Connect the 5500A TC terminals to the UUT TC INPUT 12.016 DISP using Type T thermocouple wire and connectors. 12.017 DISP Press to select an input function. 12.017 DISP Press to select thermocouple function. 12.017 DISP Select Type T thermocouple, and press . 12.017 DISP Select degrees Celsius, and press . 12.018 HEAD {INPUT(read)} 12.019 M550 TCREF 21degC 12.020 5500 -250.0degC 2U _T TC 2W 13.001 5500 -200.0degC 1U _T TC 2W 14.001 5500 -100.0degC 0.5U _T TC 2W 15.001 5500 400.0degC 0.3U _T TC 2W 16.001 M550 * 16.002 HEAD {OUTPUT(source)} 16.003 DISP Change the thermocouple connections on the UUT 16.003 DISP from TC-INPUT to TC-OUTPUT. 16.004 DISP Press to select an output function. 16.004 DISP Press to select thermocouple function. 16.004 DISP Select Type T thermocouple and press . 16.004 DISP Select degrees Celsius, and press . 16.004 DISP Press <4><0><0> to enter 400degC. 16.005 5500 400.0degC 0.3U _T TM 2W 17.001 JMP 41.001 17.002 LABEL TYPE_R 17.003 HEAD {R-type THERMOCOUPLE} 17.004 DISP Connect the 5500A TC terminals to the UUT TC INPUT 17.004 DISP using Type R thermocouple wire and connectors. 17.005 DISP Press to select an input function. 17.005 DISP Press to select thermocouple function. 17.005 DISP Select Type R thermocouple, and press . 17.005 DISP Select degrees Celsius, and press . 17.006 HEAD {INPUT(read)} 17.007 M550 TCREF 21degC # -50C to -1C Range not verified (5500A Output N/A). 17.008 5500 400degC 1U _R TC 2W 18.001 5500 1500degC 0.6U _R TC 2W 19.001 M550 * 19.002 HEAD {OUTPUT(source)} 19.003 DISP Change the thermocouple connections on the UUT 19.003 DISP from TC-INPUT to TC-OUTPUT. 19.004 DISP Press to select an output function. 19.004 DISP Press to select thermocouple function. 19.004 DISP Select Type T thermocouple and press . 19.004 DISP Select degrees Celsius, and press . 19.004 DISP Press <4><0><0> to enter 400degC. 19.005 5500 400degC 1U _R TM 2W 20.001 JMP 41.001 20.002 LABEL TYPE_S 20.003 HEAD {S-type THERMOCOUPLE} 20.004 DISP Connect the 5500A TC terminals to the UUT TC INPUT 20.004 DISP using Type S thermocouple wire and connectors. 20.005 DISP Press to select an input function. 20.005 DISP Press to select thermocouple function. 20.005 DISP Select Type S thermocouple, and press . 20.005 DISP Select degrees Celsius, and press . 20.006 HEAD {INPUT(read)} 20.007 M550 TCREF 21degC # -50C to -26C Range not verified (5500A Output N/A). 20.008 5500 400degC 1U _S TC 2W 21.001 5500 1500degC 0.7U _S TC 2W 22.001 M550 * 22.002 HEAD {OUTPUT(source)} 22.003 DISP Change the thermocouple connections on the UUT 22.003 DISP from TC-INPUT to TC-OUTPUT. 22.004 DISP Press to select an output function. 22.004 DISP Press to select thermocouple function. 22.004 DISP Select Type T thermocouple and press . 22.004 DISP Select degrees Celsius, and press . 22.004 DISP Press <4><0><0> to enter 400degC. 22.005 5500 400degC 1U _S TM 2W 23.001 JMP 41.001 23.002 LABEL TYPE_B 23.003 HEAD {B-type THERMOCOUPLE} 23.004 DISP Connect the 5500A TC terminals to the UUT TC INPUT 23.004 DISP using Type B thermocouple wire and connectors. 23.005 DISP Press to select an input function. 23.005 DISP Press to select thermocouple function. 23.005 DISP Select Type B thermocouple, and press . 23.005 DISP Select degrees Celsius, and press . 23.006 HEAD {INPUT(read)} 23.007 M550 TCREF 21degC # +100C to +139C Range not verified (5500A Output N/A) # +140C to +239C Range not verified (5500A Output N/A) # +240C to +399C Range not verified (5500A Output N/A) 23.008 5500 849degC 1U _B TC 2W 24.001 5500 1820.0degC 0.7U _B TC 2W 25.001 M550 * 25.002 HEAD {OUTPUT(source)} 25.003 DISP Change the thermocouple connections on the UUT 25.003 DISP from TC-INPUT to TC-OUTPUT. 25.004 DISP Press to select an output function. 25.004 DISP Press to select thermocouple function. 25.004 DISP Select Type T thermocouple and press . 25.004 DISP Select degrees Celsius, and press . 25.004 DISP Press <1><0><0><0> to enter 1000degC. 25.005 5500 1000.0degC 0.7U _B TM 2W 26.001 JMP 41.001 26.002 LABEL TYPE_K 26.003 HEAD {K-type THERMOCOUPLE} 26.004 DISP Connect the 5500A TC terminals to the UUT TC INPUT 26.004 DISP using Type K thermocouple wire and connectors. 26.005 DISP Press to select an input function. 26.005 DISP Press to select thermocouple function. 26.005 DISP Select Type K thermocouple, and press . 26.005 DISP Select degrees Celsius, and press . 26.006 HEAD {INPUT(read)} 26.007 M550 TCREF 21degC # -240C to -201C Range not verified (5500A Output N/A). 26.008 5500 -150.0degC 1U _K TC 2W 27.001 5500 1000.0degC 0.5U _K TC 2W 28.001 M550 * 28.002 HEAD {OUTPUT(source)} 28.003 DISP Change the thermocouple connections on the UUT 28.003 DISP from TC-INPUT to TC-OUTPUT. 28.004 DISP Press to select an output function. 28.004 DISP Press to select thermocouple function. 28.004 DISP Select Type T thermocouple and press . 28.004 DISP Select degrees Celsius, and press . 28.004 DISP Press <1><0><0><0> to enter 1000degC. 28.005 5500 1000.0degC 0.5U _K TM 2W 29.001 JMP 41.001 29.002 LABEL TYPE_J 29.003 HEAD {J-type THERMOCOUPLE} 29.004 DISP Connect the 5500A TC terminals to the UUT TC INPUT 29.004 DISP using Type J thermocouple wire and connectors. 29.005 DISP Press to select an input function. 29.005 DISP Press to select thermocouple function. 29.005 DISP Select Type J thermocouple, and press . 29.005 DISP Select degrees Celsius, and press . 29.006 HEAD {INPUT(read)} 29.007 M550 TCREF 21degC 29.008 5500 -200.0degC 0.9U _J TC 2W 30.001 5500 -140.0degC 0.5U _J TC 2W 31.001 5500 1000.0degC 0.3U _J TC 2W 32.001 M550 * 32.002 HEAD {OUTPUT(source)} 32.003 DISP Change the thermocouple connections on the UUT 32.003 DISP from TC-INPUT to TC-OUTPUT. 32.004 DISP Press to select an output function. 32.004 DISP Press to select thermocouple function. 32.004 DISP Select Type T thermocouple and press . 32.004 DISP Select degrees Celsius, and press . 32.004 DISP Press <1><0><0><0> to enter 1000degC. 32.005 5500 1000.0degC 0.3U _J TM 2W 33.001 JMP 41.001 33.002 LABEL TYPE_N 33.003 HEAD {N-type THERMOCOUPLE} 33.004 DISP Connect the 5500A TC terminals to the UUT TC INPUT 33.004 DISP using Type N thermocouple wire and connectors. 33.005 DISP Press to select an input function. 33.005 DISP Press to select thermocouple function. 33.005 DISP Select Type N thermocouple, and press . 33.005 DISP Select degrees Celsius, and press . 33.006 HEAD {INPUT(read)} 33.007 M550 TCREF 21degC # -250C to -231C Range not verified (5500A Output N/A). # -230C to -201C Range not verified (5500A Output N/A). 33.008 5500 -100.0degC 1U _N TC 2W 34.001 5500 1000.0degC 0.4U _N TC 2W 35.001 M550 * 35.002 HEAD {OUTPUT(source)} 35.003 DISP Change the thermocouple connections on the UUT 35.003 DISP from TC-INPUT to TC-OUTPUT. 35.004 DISP Press to select an output function. 35.004 DISP Press to select thermocouple function. 35.004 DISP Select Type T thermocouple and press . 35.004 DISP Select degrees Celsius, and press . 35.004 DISP Press <5><0><0> to enter 500degC. 35.005 5500 500.0degC 0.4U _N TM 2W 36.001 JMP 41.001 36.002 LABEL TYPE_E 36.003 HEAD {E-type THERMOCOUPLE} 36.004 DISP Connect the 5500A TC terminals to the UUT TC INPUT 36.004 DISP using Type E thermocouple wire and connectors. 36.005 DISP Press to select an input function. 36.005 DISP Press to select thermocouple function. 36.005 DISP Select Type E thermocouple, and press . 36.005 DISP Select degrees Celsius, and press . 36.006 HEAD {INPUT(read)} 36.007 M550 TCREF 21degC 36.008 5500 -230.0degC 2U _E TC 2W 37.001 5500 -200.0degC 1U _E TC 2W 38.001 5500 -150.0degC 0.5U _E TC 2W 39.001 5500 1000.0degC 0.3U _E TC 2W 40.001 M550 * 40.002 HEAD {OUTPUT(source)} 40.003 DISP Change the thermocouple connections on the UUT 40.003 DISP from TC-INPUT to TC-OUTPUT. 40.004 DISP Press to select an output function. 40.004 DISP Press to select thermocouple function. 40.004 DISP Select Type T thermocouple and press . 40.004 DISP Select degrees Celsius, and press . 40.004 DISP Press <1><0><0><0> to enter 1000degC. 40.005 5500 1000.0degC 0.3U _E TM 2W 41.001 END